DatasheetView PDF DatasheetPrinter Friendly DatasheetPrint This DatasheetAdd to Compare ListHave Questions? Ask UsRequest for a Quotation today

SCT elisa kit :: General Salmon Calcitonin (SCT) ELISA Kit

Scan QR to view Datasheet Catalog #    MBS2000203 (SPECIAL PROMOTION: Get FREE Starbucks gift card ($5 for 48-Wells-Kit or $10 for 96-Wells-Kit. Maximum 5 gift cards/product). Gift card(s) will be included with shipment)
Typical Testing Data/Standard Curve (for reference only)
Unit / Price
24-Strip-Wells  /  $305 +1 FREE 8GB USB
48-Strip-Wells  /  $630 +1 FREE 8GB USB
96-Strip-Wells  /  $855 +1 FREE 8GB USB
5x96-Strip-Wells  /  $3,525 +4 FREE 8GB USB
10x96-Strip-Wells  /  $6,395 +7 FREE 8GB USB
 
 Product Name   

Salmon Calcitonin (SCT), ELISA Kit

 Also Known As   

Salmon Calcitonin (SCT) ELISA Kit

 Product Synonym Names    CGRP; Salcatonin
 Product Gene Name   

SCT elisa kit

[Similar Products]
 Research Use Only    For Research Use Only. Not for use in diagnostic procedures.
Table BarTOPTable Bar
  Manual Insert    Download PDF Manual #PDF_Manual" title="View Salmon Calcitonin (SCT) ELISA Kit (MBS2000203) PDF Manual">View PDF Manual
 Species Reactivity    General
Section Bar
 Specificity    This assay has high sensitivity and excellent specificity for detection of Salmon Calcitonin (SCT).
No significant cross-reactivity or interference between Salmon Calcitonin (SCT) and analogues was observed.
Section Bar
 Samples    Serum, plasma and other biological fluids
 Assay Type    Competitive
 Detection Range    246.91-20000pg/mL
 Sensitivity    Typically less than 89.4pg/mL.
 Precision    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Salmon Calcitonin (SCT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Salmon Calcitonin (SCT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
 Assay Procedure Summary    1. Prepare all reagents, samples and standards;
2. Add 50uL standard or sample to each well.
    And then add 50uL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37 degree C;
3. Aspirate and wash 3 times;
4. Add 100uL prepared Detection Reagent B. Incubate 30 minutes at 37 degree C;
5. Aspirate and wash 5 times;
6. Add 90uL Substrate Solution. Incubate 15-25 minutes at 37 degree C;
7. Add 50uL Stop Solution. Read at 450 nm immediately.
Section Bar
 Preparation and Storage    For unopened kit: All the reagents should be kept according to the labels on vials. The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 degree C upon receipt while the others should be at 4 degree C.

For opened kit: When the kit is opened, the remaining reagents still need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and reseal along entire edge of zip-seal.

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
 Other Notes    Small volumes of SCT elisa kit vial(s) may occasionally become entrapped in the seal of the product vial during shipment and storage. If necessary, briefly centrifuge the vial on a tabletop centrifuge to dislodge any liquid in the container`s cap. Certain products may require to ship with dry ice and additional dry ice fee may apply.
 Searchable Terms for SCT purchase    MBS2000203 is a ready-to-use microwell, strip plate Competitive ELISA (enzyme-linked immunosorbent assay) Kit for analyzing the presence of the Salmon Calcitonin (SCT) ELISA Kit target analytes in biological samples. The concentration gradients of the kit standards or positive controls render a theoretical kit detection range of 246.91-20000pg/mL in biological research samples containing SCT, with an estimated sensitivity of Typically less than 89.4pg/mL. The ELISA analytical biochemical technique of the MBS2000203 kit is based on SCT antibody-SCT antigen interactions (immunosorbency) and an HRP colorimetric detection system to detect SCT antigen targets in samples. The ELISA Kit is designed to detect native, not recombinant, SCT. Appropriate sample types may include undiluted human body fluids and/or tissue homogenates, secretions such as Serum, plasma and other biological fluids. Quality control assays assessing reproducibility identified the intra-assay CV (%) and inter-assay CV(%).
Table BarTOPTable Bar
 

Product Description specifically for SCT elisa kit

   Principle of the Assay: This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Salmon Calcitonin (SCT) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Salmon Calcitonin (SCT) and unlabeled Salmon Calcitonin (SCT) (Standards or samples) with the pre-coated antibody specific to Salmon Calcitonin (SCT). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Salmon Calcitonin (SCT) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Salmon Calcitonin (SCT) in the sample.
Section Bar
 Typical Testing Data/Standard Curve (for reference only) of SCT elisa kit    SCT elisa kit Typical Testing Data/Standard Curve (for reference only) image
Table BarTOPTable Bar
 Protein Family    Secretin
 Precautions    All of MyBioSource's Products are for scientific laboratory research purposes and are not for diagnostic, therapeutics, prophylactic or in vivo use. Through your purchase, you expressly represent and warrant to MyBioSource that you will properly test and use any Products purchased from MyBioSource in accordance with industry standards. MyBioSource and its authorized distributors reserve the right to refuse to process any order where we reasonably believe that the intended use will fall outside of our acceptable guidelines.
 Disclaimer    While every efforts were made to ensure the accuracy of the information provided in this datasheet, MyBioSource will not be liable for any omissions or errors contained herein. MyBioSource reserves the right to make changes to this datasheet at any time without prior notice.

It is the responsibility of the customer to report product performance issues to MyBioSource within 30 days of receipt of the product. Please visit our Terms & Conditions page for more information.
Table BarTOPTable Bar
horiz bar
 SSL   Follow us on Facebook Follow us onTwitter Follow us on Google Plus Connect us on LinkedIn Subscribe to our RSS Feed for latest products and special promotions